Customization: | Available |
---|---|
After-sales Service: | Support Operation Guidance |
Power Supply: | 380V |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Relevant standard test requirements
Meet the standards: IEC.62133-2012,UL.1642, United Nations38.3 And 31241-2014 standard
Temperature cycle test chamber: Put the battery according to 4.5.1 After the specified test method is fully charged, place the battery in the test chamber and follow the following steps to test (see Figure 1)):
A) Maintain 6h at a temperature of 75ºC±2ºC;
B) Maintain 6h at a temperature of -40ºC±2ºC;
C) Repeat step a)~b), a total of 10 cycles;
D) Restore the ambient temperature of 20ºC±5ºC.
During the test, the conversion time between each of the two temperatures is not more than 30min.
Main parameters
Model | MHP-150 | MHP-225 | MHP-408 | MHP-800 | MHP-1000 | |||||||||||||||||||||||||||||||||||||
Inner box size W*H*D(cm) | 50*60*50 | 60*75*50 | 80*85*60 | 100*100*80 | 100*100*100 | |||||||||||||||||||||||||||||||||||||
Outer box size W*H*D(cm) | 70*150*165 | 80*165*170 | 100*175*180 | 120*190*185 | 120*190*190 | |||||||||||||||||||||||||||||||||||||
The volume of the inner box | 150L | 225L | 408L | 800L | 1000L | |||||||||||||||||||||||||||||||||||||
Performance | Temperature and humidity range | 0ºC/20°ºC/-40°C/-60C/-70ºC/+100ºC(+150ºC): 10%RH-98% | ||||||||||||||||||||||||||||||||||||||||
Volatility of temperature and humidity | ±0.3ºC-±0.5ºC; ±2.5%RH | |||||||||||||||||||||||||||||||||||||||||
Temperature and humidity deviation | ±2ºC; Soil 3RH(>75%RH); Soil 5%RH ≤75 % RH) | |||||||||||||||||||||||||||||||||||||||||
Controller analysis accuracy | ±0.01ºC; ±0.1%RH | |||||||||||||||||||||||||||||||||||||||||
Warm-up time | 100 | 150 | 100 | 150 | 100 | 150 | 100 | 150 | 100 | 150 | ||||||||||||||||||||||||||||||||
(RT-HT) | 30 | 40 | 40 | 50 | 45 | 55 | 55 | 60 | 50 | 60 | ||||||||||||||||||||||||||||||||
Cooling time | 0 | -20 | -40 | -60 | -70 | 0 | -20 | -40 | -60 | -10 | 0 | -20 | -40 | -60 | -70 | 0 | -20 | 40 | -60 | -70 | 0 | -20 | -40 | -60 | -70 | |||||||||||||||||
RT.-HT. | 40 | 50 | 55 | 90 | 100 | 40 | 50 | 55 | 90 | 100 | 40 | 50 | 55 | 90 | 100 | 40 | 50 | 55 | 90 | 100 | 40 | 50 | 55 | 90 | 100 | |||||||||||||||||
Power supply (K.w.) | 3.0 | 4.0 |
6.0 | 6.0 | 6.0 | 3.5 | 5.5 | 6.5 | 8.0 | 8.5 | 4.5 | 6.0 | 7.5 | 9.5 | 9.5 | 5.5 | 9.0 | Ten | 10.5 | 10.5 | 10.5 | 10.5 | 10.5 | 10.5 | 10.5 | |||||||||||||||||
Weight (Kg) | 250 | 300 | 400 | 600 | 700 | |||||||||||||||||||||||||||||||||||||
Material | Inner wall | SUS#304 stainless steel 2B panel | ||||||||||||||||||||||||||||||||||||||||
Outer casing | High-strength cold-rolled steel plate double-sided fine powder baking paint | |||||||||||||||||||||||||||||||||||||||||
System wind circuit | Glass fiber + polyurethane foam | |||||||||||||||||||||||||||||||||||||||||
System | Way | Centrifugal fan one broadband forced airflow circulation (top out and down) | ||||||||||||||||||||||||||||||||||||||||
Refrigeration method | Single-stage mechanical compression refrigeration (or binary stacked refrigeration) | |||||||||||||||||||||||||||||||||||||||||
Chiller | Single fully enclosed piston type French Taikang compressor | |||||||||||||||||||||||||||||||||||||||||
Refrigerant | R404A American Shebang Environmental Protection Cold Coal (R23+404A) | |||||||||||||||||||||||||||||||||||||||||
Condensation method | Air-cooled or water-cooled | |||||||||||||||||||||||||||||||||||||||||
Heater | Alloy electric heating wire heater | |||||||||||||||||||||||||||||||||||||||||
Humidifier | Semi-closed steam humidification (boiler type) | |||||||||||||||||||||||||||||||||||||||||
Water supply method | For fully automatic circulating water supply (water quality requires resistivity>500 M) | |||||||||||||||||||||||||||||||||||||||||
Standard configuration | 1 observation window (double-layer insulated tempered glass); 1 test hole 50mm (located on the left); 2-layer sample rack; 1 box lighting fixture (fluorescent lamp); 1 pack of gauze; 8 hooks; 2 fuses; 4 small wheels; 1 power cord | |||||||||||||||||||||||||||||||||||||||||
SafeProtect | Equipped with an explosion-proof pressure relief port, the box door is equipped with an explosion-proof chain, and the observation window is equipped with an explosion-proof grid. Optional smoke alarm, smoke exhaust pipe, video surveillance and fire extinguishing system. |